Troubleshooting

Problem Possible Cause Solution

Calibration Curve

• 1000 ng/mL Calibrator
(calculated) is out of range

• 100 ng/mL Calibrator
(calculated) is out of range

Correlation coefficient
(r2) is out of range

Improper identification of Calibrator wells in Calculator Correct numeric value(s) and recalculate the Calibration Curve

Calibrators not mixed thoroughly

Repeat Assay Part 3 using new Calibrator Panel
Cross-contamination between wells Repeat Assay Part 3
Deterioration of Calibrators Repeat Assay Part 3 using new Calibrator Panel
Contamination of Calibrators Repeat Assay Part 3 using new Calibrator Panel
Improper volume of ATP Calibrator dispensed into wells

Luminometer related problem
Repeat Assay Part 3


Refer to luminometer manufacturer’s recommendations; freeze Assay Plate strips

Non-Stimulated Control Specimen

• ATP result is >60 ng/mL

Blood was diluted improperly Refer to procedure for proper protocol; repeat test using fresh specimens
An anti-coagulant other than
sodium heparin was used
Repeat test using fresh specimens
collected into a sodium heparin tube
Cross contamination of wells Decrease setting of plate shaker to allow vigorous shaking without splashing
Calibration Curve is invalid Refer to first section of this table
Improper volume of lysate transferred into Measurement Plate Repeat Assay Part 3
Improper volume of Calibrators dispensed into Measurement Plate Repeat Assay Part 3
Deterioration of Calibrators Repeat Assay Part 3 using new Calibrator Panel
The ATP value may be accurate for the individual chosen as the control Choose another source for control specimen and repeat test using fresh
specimens

Stimulated Control Specimen

• ATP result is <240 ng/mL

Blood specimen is not fresh Repeat test using fresh specimens
Blood specimen was subjected to temperature changes or excessive mixing Repeat test using fresh specimens
An anti-coagulant other than sodium heparin was used Repeat test using fresh specimens collected into a sodium heparin tube
Blood was diluted improperly Refer to procedure for proper protocol; repeat test using fresh specimens
Activity of Stimulant has decreased Repeat test using fresh specimens and use a new vial of Stimulant
Improper volumes of specimen, Stimulant, or other reagents were used Refer to procedure for proper protocol; repeat test using fresh specimens
Calibration curve is invalid Refer to first section of this table
Improper volume of lysate transferred into Measurement Plate Repeat Assay Part 3
Improper volume of Calibrators dispensed into Measurement Plate Repeat Assay Part 3
Vacuum strength is too high, causing loss of Dynabeads® Adjust the level of the vacuum strength to achieve a slow, gentle aspiration
Plate shaker is not vigorous enough for optimal mixing Increase setting of plate shaker to allow vigorous shaking without splashing
Incubation time <15 hours Refer to procedure for proper protocol; repeat test using fresh specimens
Incubator temperature and/or CO2 content is out of range Adjust incubator to 37C / 5% CO2 and repeat the test using fresh specimens
The ATP value may be representative for the individual chosen as the control Choose another source for control specimen and repeat test using fresh specimens